NEW: the ISO 9001:2026, ISO 14001:2026, ISO/IEC 17020:2026 and FSSC 22000 V7 documentation kits are now available.
ISO/IEC 17025

ISO/IEC 17025:2017 Method Validation: Purpose, Extent and Records

Published on July 17, 2019
12 min read
By QSE Academy Team

Last Updated on November 25, 2026 by Hafsa J.

Follow QSE Academy on Google

Add us as a preferred source to see our ISO and compliance articles more often in your Google results and Discover.

QSE Academy Team

QSE Academy Team

Clause 3.9 of ISO/IEC 17025:2017 defines validation in eleven words: verification, where the specified requirements are adequate for an intended use. The definition is short, and it already contains the whole subject.

Method validation is not a heavier kind of verification. Verification asks whether your lab can run a method as specified. Validation asks whether the method suits the use you intend to make of it. The standard requires the second only in defined cases, and it does not require more validation than the application needs. Here is when to validate, how far to go, and what to keep. Have a method that sits between the two? Describe it in the comments and we will give you our reading.

Happy reading.

 

ISO/IEC 17025:2017 method validation in 30 seconds

Verification is for every methodBefore introducing a method, the lab verifies that it can achieve the required performance and keeps the records (7.2.1.5).
Validation is for three casesNon-standard methods, laboratory-developed methods, and standard methods used outside their intended scope or otherwise modified (7.2.2.1).
The extent is set by the useAs extensive as is necessary to meet the needs of the given application or field of application.
Five records close the fileClause 7.2.2.4 lists them, ending with a statement on the validity of the method for its intended use.

Two definitions that settle most arguments

Both terms are defined in clause 3, and both definitions come from ISO/IEC Guide 99, the international vocabulary of metrology. Verification (3.8) is the provision of objective evidence that a given item fulfils specified requirements. Validation (3.9) is a verification where the specified requirements are adequate for an intended use.

The difference sits in the last three words. A verification compares an item to requirements that someone has already written. A validation also has to show that those requirements are the right ones for the job. Note 5 to the definition of verification says it in one line: not every verification is a validation.

The example attached to clause 3.9 makes the point with one change of matrix. A measurement procedure ordinarily used for the mass concentration of nitrogen in water may be validated also for the measurement of nitrogen in human serum. The procedure is the same. The intended use is new, so the evidence has to be produced again for that use.

Verify or validate: which one applies to your method?

Clause 7.2 is split along the same line. Clause 7.2.1 deals with the selection and verification of methods, clause 7.2.2 with their validation. The origin of the method decides which one applies.

What clause 7.2 asks, by origin of the method

ISO/IEC 17025:2017, clauses 7.2.1.5 and 7.2.2.1

Standard method, used as published

Verify

The lab verifies that it can properly perform the method before introducing it, by showing that it can achieve the required performance (7.2.1.5).

Non-standard or laboratory-developed method

Validate

The first two cases named in clause 7.2.2.1.

Standard method outside its intended scope, or otherwise modified

Validate

The third case of clause 7.2.2.1: the published method no longer covers what the lab does with it.

So what?A published standard is not a permanent exemption. The moment a lab changes the matrix, the range or a step of the method, it has moved from the first card to the third.

Source: ISO/IEC 17025:2017, clauses 7.2.1.5 and 7.2.2.1.

 

What verification requires

Clause 7.2.1.5 has three sentences. The lab verifies that it can properly perform methods before introducing them, by making sure it can achieve the required performance. Records of the verification are retained. If the method is revised by the issuing body, verification is repeated to the extent necessary.

The third sentence is the one that gets lost. A new edition of a published method does not always mean a full new verification. It always means a decision about how much of the verification has to be repeated, and that decision is easier to defend when it is written down.

Where the methods come from

When the customer does not specify the method, the lab selects an appropriate one and informs the customer of its choice (7.2.1.4). The same clause recommends methods published in international, regional or national standards, by reputable technical organizations, in relevant scientific texts or journals, or as specified by the manufacturer of the equipment. It adds that laboratory-developed or modified methods can also be used. The standard recommends published methods and does not exclude an in-house one. It asks for a different kind of evidence.

When a method has to be developed, clause 7.2.1.6 makes it a planned activity, assigned to competent personnel with adequate resources, with periodic review to confirm that the needs of the customer are still being fulfilled. Modifications to the development plan are approved and authorized.

How far does an ISO/IEC 17025:2017 validation have to go?

Clause 7.2.2.1 answers in one sentence: the validation is as extensive as is necessary to meet the needs of the given application or field of application. No list of mandatory experiments follows. A note names the techniques that can be used, alone or in combination.

a) Reference standards or materials

Calibration or evaluation of bias and precision using reference standards or reference materials.

b) Influencing factors

Systematic assessment of the factors influencing the result.

c) Robustness testing

Variation of controlled parameters, such as incubator temperature or volume dispensed.

d) Comparison of methods

Comparison of results achieved with other validated methods.

e) Interlaboratory comparisons

Results set against those of other laboratories.

f) Uncertainty evaluation

Evaluation of measurement uncertainty based on the theoretical principles of the method and practical experience of its performance.

Another note widens the perimeter: validation can include procedures for sampling, handling and transportation of test or calibration items. Both are notes, so they describe options. The requirement is the sentence above them, and it cuts both ways. A minor extension of range does not call for the same amount of work as a method the lab wrote from nothing. The application decides, and the lab has to be able to justify the extent it chose.

The performance characteristics to choose from

Clause 7.2.2.3 requires that the performance characteristics of validated methods, as assessed for the intended use, be relevant to the customers’ needs and consistent with specified requirements. Its note says they can include, but are not limited to, the characteristics in the table.

Characteristic named in the note to 7.2.2.3The question it answers, in plain words
Measurement range and linearityOver which interval does the method respond as expected?
Accuracy and biasHow close are results to a reference value?
Repeatability or reproducibilityHow much do results scatter when the measurement is repeated?
Limit of detection and limit of quantificationWhat is the smallest amount the method can detect, then quantify?
Selectivity of the methodDoes the method respond to the intended analyte or property?
Robustness against external influences, or cross-sensitivity against interference from the matrixWhat happens when conditions or the sample matrix vary?
Measurement uncertainty of the resultsWhat interval goes with a reported value?

The left column groups the terms of the note. The right column turns each one into a plain question. The note does not make every characteristic mandatory for every method: a qualitative test has no linearity to study. What the clause requires is that the characteristics you assess are the ones your customer’s need calls for, and that you can say why. The link with uncertainty is direct: the note names the measurement uncertainty of the results among the characteristics, and the budget behind it is described on our page about uncertainty in measurement.

The five records of a validation file (7.2.2.4)

Clause 7.2.2.4 lists the records of validation the lab retains. Read in order, the five items are also a table of contents for the file.

1
The validation procedure used

Item a). How the validation was carried out: the techniques chosen and the way the experiments were run.

2
Specification of the requirements

Item b). What the method has to achieve for its intended use. Without this record, results have nothing to be compared to.

3
Determination of the performance characteristics

Item c). The characteristics that were assessed for the method.

4
Results obtained

Item d). The data and what was calculated from them.

5
A statement on the validity of the method

Item e). The statement details the method’s fitness for the intended use. It is the conclusion of the file.

One other clause decides who may do this work and sign it. Clause 6.2.6 a) requires the lab to authorize personnel for the development, modification, verification and validation of methods. A validation statement signed by someone with no such authorization is a gap in clause 6.2, even when the experiments are sound.

The order of items b) and d) carries a practical lesson that the standard does not spell out but that follows from the definition in 3.9. A validation shows that requirements adequate for the intended use are fulfilled. Requirements written after the results are known cannot play that role. To prepare the plan before the bench work, the ISO/IEC 17025 Method Validation Plan Generator is available as an online tool.

After validation: changes and deviations

A validated method does not stay validated whatever happens to it. Under clause 7.2.2.2, when changes are made to a validated method, their influence is determined, and where they are found to affect the original validation, a new method validation is performed. The clause asks for two steps: first an assessment of the change, then a new validation only if the assessment shows an effect.

A deviation is a different animal. Clause 7.2.1.7 allows deviations from methods only if the deviation has been documented, technically justified, authorized and accepted by the customer. The note adds that customer acceptance can be agreed in advance in the contract. A change is permanent and belongs to the method. A deviation is an exception on a given job, and it shows up on the report, since clause 7.8.2.1 n) lists additions to, deviations or exclusions from the method among the report items.

Methods also have to stay current as documents. Clause 7.2.1.2 requires methods, procedures and supporting documentation to be kept up to date and readily available to personnel, with a cross-reference to the control of documents in 8.3, and clause 7.2.1.3 requires the latest valid version of a method unless it is not appropriate or possible to use it. Document control belongs to the management system, the area of our page on ISO/IEC 17025 clause 8 management system options.

Two FDA texts that use the same split

US food and drug regulation gives two public examples of the verification and validation pair at work.

The first is a regulation. For laboratories accredited under the FDA program for analyses of foods, 21 CFR 1.1151 requires the method of analysis to be fit for purpose, within the lab’s scope of LAAF-accreditation, appropriately validated, and appropriately verified by the laboratory. For a verification, the same section lists what is recorded: the method, the intended purpose of the analysis, the results of the verification, the procedure used, supporting analytical data, and whether the laboratory is able to properly perform the method.

The second is a laboratory manual that FDA publishes. The ORA Laboratory Manual, Volume II, section 5.4.5, on methods, method verification and validation (revision 03, revised October 31, 2023), states that standard or FDA official methods require verification as confirmation that the laboratory can meet the test method performance specifications, and that this verification must be completed before the method is used for routine testing. Its stated purpose is to provide basic requirements to the laboratories of the FDA Office of Regulatory Science. It is an illustration of the same logic at a regulator, not a requirement for other labs.

Where method validation sits in an accreditation project

Validation files take bench time, and bench time has to be planned alongside everything else a lab prepares before an assessment. Planning is the topic of our page on the ISO/IEC 17025 accreditation timeline. Clause 7.2 is also one part of a longer chain that starts with the review of requests, where clause 7.1.1 d) asks that the methods selected be capable of meeting the customers’ requirements. The chain is the subject of our page on the process requirements of ISO 17025.

For the written side, the QSE Academy ISO/IEC 17025:2017 documentation kit lists a Method Selection Procedure, a Method Validation and Verification Procedure, a Standard Method Validation File and a Register of Method Validations and Revalidations.

Questions about ISO/IEC 17025:2017 method validation

Does a standard method need to be validated?+

Not when it is used as published and within its intended scope: clause 7.2.1.5 requires verification. Clause 7.2.2.1 requires validation when a standard method is used outside its intended scope or otherwise modified.

Is method validation required for calibration methods too?+

Yes. Clause 7.2 applies to methods for all laboratory activities, and the note to 7.2.2.1 mentions test or calibration items.

Does a change to a validated method always require a new validation?+

No. Clause 7.2.2.2 requires the influence of the change to be determined. A new validation is performed where the change is found to affect the original validation.

Must a published standard be rewritten as an internal procedure?+

The note to 7.2.1.3 says standards that contain sufficient and concise information, written so that operating personnel can use them, do not need to be supplemented or rewritten as internal procedures. Additional documentation can be necessary for optional steps.

What is the last document in a validation file?+

Clause 7.2.2.4 e): a statement on the validity of the method, detailing its fitness for the intended use.

Validation, restated: evidence that the requirements you set are adequate for what you intend to do with the method, and that the method meets them.

Go further

The clause 7 documents of the ISO/IEC 17025:2017 documentation kit include a Method Validation and Verification Procedure and a Standard Method Validation File.

Sources

  • ISO/IEC 17025:2017, General requirements for the competence of testing and calibration laboratories, third edition, November 2017: clauses 3.8, 3.9, 6.2.6, 7.1.1, 7.2.1, 7.2.2 and 7.8.2.1.
  • The FDA requirements for methods of analysis used by LAAF-accredited laboratories: 21 CFR 1.1151, Electronic Code of Federal Regulations.
  • FDA, Office of Regulatory Affairs, document MAN-000037, revision 03 of October 31, 2023: ORA Lab Manual Vol. II, Methods, Method Verification and Validation (ORA-LAB.5.4.5).
QSE Academy Team

The QSE Academy Team writes QSE Academy's guides on information security, IT service management and the other technical standards we cover.
Every article is checked against the text of the standard it covers before publication, and its sources are listed at the end.

Take the next step

Prepare your ISO/IEC 17025 certification with the complete kit

The complete documentation package to deploy an ISO/IEC 17025:2017-compliant laboratory management system: ready-to-use document architecture covering competence, methods, metrological traceability, uncertainty, and statements of conformity.

Leave your thought here

Your email address will not be published. Required fields are marked *