Last Updated on November 27, 2026 by QSE Academy Team
Follow QSE Academy on Google
Add us as a preferred source to see our ISO and compliance articles more often in your Google results and Discover.
On November 25, 2026, the FDA issued an Early Alert about certain lots of Portex endotracheal tubes made by ICU Medical. The company traced the problem to a resin with variable plasticizer content that can make neonatal and infant tubes softer than expected, and has reported one death associated with the issue.
ICU Medical Portex alert: the essentials in 30 seconds
The action
An FDA Early Alert on a device removal. ICU Medical told customers on November 20 to stop using, then destroy or quarantine the affected lots.
The products
Five uncuffed tracheal tube item numbers, 2.5 mm to 3.5 mm, in the Murphy eye and Polar preformed ranges.
The cause
A resin with variability in the percentage of plasticizer, which may leave the tube material softer than expected.
The ISO 13485 link
Clauses 7.4.2 and 7.4.3 on purchasing information and verification of purchased product, and 8.3.3 on advisory notices.
The facts
On November 25, 2026, the FDA Center for Devices and Radiological Health posted an Early Alert on its recalls and early alerts page about ICU Medical Portex endotracheal tubes to notify the public of a potentially high-risk device issue. The agency lists the status as information still being collected and says it will update the page as significant new information becomes available.
According to the FDA, ICU Medical sent a letter to affected customers on November 20, 2026. It asks them to check every inventory location, stop using the affected tubes immediately and destroy them, or segregate and quarantine them when destruction is not possible right away. The letter also asks customers who distributed the tubes to forward the notice and its response form to their own customers.
The alert lists five item numbers: uncuffed Murphy eye oral and nasal tubes in 2.5, 3.0 and 3.5 mm, and uncuffed Polar oral tubes in 3.0 and 3.5 mm, with a full list of affected lots. As of November 16, 2026, ICU Medical had reported one death associated with the issue. In Canada, Health Canada published a Type I recall of the same five catalog numbers on November 19, 2026, describing a lot-specific manufacturing issue.
The quality failure identified
ICU Medical identified a manufacturing issue affecting certain lots: the tubes were made with a resin whose percentage of plasticizer varies, which may make the tube material softer than expected. The FDA explains the clinical consequence for the neonatal and early infant population: a delay in intubation due to increased resistance to ventilation, tube kinking, and difficulty intubating because of the soft material. If the problem is not spotted during intubation, it may lead to inadequate ventilation and unscheduled tube replacement, which may contribute to serious adverse events, including death.
Since February 2, 2026, the FDA Quality Management System Regulation (QMSR) incorporates ISO 13485:2016 by reference into 21 CFR Part 820, so the standard’s clauses now frame what FDA investigators expect from US manufacturers.
Clause 7.4.2 requires purchasing information to describe product specifications and requirements for product acceptance. Clause 7.4.3 requires inspection or other verification activities whose extent is proportionate to the risks of the purchased product. Once product is in the field, clause 8.3.3 requires action appropriate to the effects of the nonconformity and documented procedures for issuing advisory notices, capable of being put into effect at any time.
What this means for quality managers
Start with your material specifications. For every polymer or compound that drives device performance, check that the purchasing documents name the property that matters, such as plasticizer content or hardness, with a tolerance and an acceptance method, not only a grade name.
Next, look at incoming verification. A certificate of analysis may be enough for a low-risk material, but clause 7.4.3 ties the extent of verification to risk. A material used in a neonatal airway device deserves a harder look at what is actually tested on each lot.
Then test your traceability. Pick one finished lot and time how long it takes to identify the resin lot behind it, and every other finished lot made from that resin. Clause 7.5.9.1 requires documented traceability procedures that define the extent of traceability; fast, lot-level traceability back to the raw material keeps any field action as narrow as possible.
Finally, rehearse your advisory notice procedure. Check that it covers inventory checks, quarantine, destruction, forwarding to downstream customers and response forms, and that complaint handling under clause 8.2.2, fed into the analysis of data required by clause 8.4, would flag reports of kinking or difficult insertion early.
Go further
To document purchasing controls, verification of purchased product, traceability and advisory notices, the ISO 13485 2016 Documentation Kit provides 106 editable Word and Excel documents organized by clause.
Sources
- U.S. Food and Drug Administration, Early Alert: ICU Medical Removes Portex Endotracheal Tubes, November 25, 2026
- Health Canada, Health product recall: Portex Endotracheal Tubes, RA-82977, November 19, 2026
- U.S. Food and Drug Administration, Quality Management System Regulation (QMSR), 21 CFR Part 820, effective February 2, 2026
- ISO 13485:2016, Medical devices, Quality management systems, Requirements for regulatory purposes, clauses 7.4.2, 7.4.3, 7.5.9, 8.2.2, 8.3.3 and 8.4
The QSE Academy Team writes QSE Academy's guides on information security, IT service management and the other technical standards we cover.
Every article is checked against the text of the standard it covers before publication, and its sources are listed at the end.